HOT SHEARS (TM) MONOPOLARCURVED SCISSORS
Report
- Report Number
- 2955842-2013-00422
- Event Type
- Malfunction
- Date Received
- February 6, 2013
- Date of Event
- December 11, 2012
- Report Date
- January 10, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Removal / Correction Number
- 2955842-051613-005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER
Narratives
THIS MDR IS BEING SUBMITTED FOR RETROSPECTIVE ACTIVITY PERFORMED RELATING TO FIELD ACTION NUMBER 2955842-051613-005 TO INVESTIGATE MICRO-CRACKS ON THE MONOPOLAR CURVED SCISSORS INSTRUMENT. THESE TYPES OF MICRO-CRACKS WILL NOT LEAD TO MECHANICAL FAILURE OF THE INSTRUMENT; HOWEVER, THERE IS A POTENTIAL FOR INSULATION FAILURE AFTER REPROCESSING, RESULTING IN A PATHWAY FOR ELECTROSURGICAL ENERGY TO LEAK TO TISSUE AND POTENTIALLY CAUSE UNINTENDED INJURIES. THE LOCATION OF THESES MICRO-CRACKS IS CONFINED TO 2 CM OF THE DISTAL END OF THE INSTRUMENT SHAFT. THIS INSTRUMENT WAS INSPECTED AS PART OF THIS RETROSPECTIVE ACTIVITY AND FOUND TO CONTAIN CRACKS ON THE INSTRUMENT MAIN TUBE.
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING PLACED THE INSTRUMENT ON AN IN-HOUSE SYSTEM FOR THE LATEX CUT TEST. SCISSORS DO NOT CUT CLEANLY THROUGH .006 LATEX. LATEX GETS SNAGGED AT SCISSOR TIPS. BLADE EDGES ARE NOT DAMAGED, BUT EDGES EXHIBIT WEAR AT THE TIPS, NEGATIVELY AFFECTING CUT PERFORMANCE. ELECTRICAL CONTINUITY PASSED. THE INSTRUMENT HAS 3 USES REMAINING. ADDITIONAL FINDINGS REVEALED THAT THE ORANGE PRINT FROM THE EXTENSION TUBE IS REMOVED. THE MISSING PRINT MEASURED ROUGHLY ABOUT 0.157 X 0.127. EVIDENCE NOT CONCLUSIVE BUT DAMAGE MOST LIKELY DUE TO IMPROPER CLEANING. PER THE CUSTOMER, AT THE TIME OF THE REPORTED EVENT, NOTHING FELL INTO A PATIENT DURING A SURGICAL PROCEDURE. THE CLEANING AND STERILIZATION USER MANUAL SPECIFICALLY STATES: WHEN SCRUBBING THE TIP OF CAUTERY INSTRUMENTS, TAKE CARE NOT TO DAMAGE THE INSULATION.
IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE, THE MONOPOLAR CURVED SCISSORS INSTRUMENT WAS DULL. THE INTENDED PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 51191 | HOT SHEARS (TM) MONOPOLARCURVED SCISSORS | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-10 | M11121016 254 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |