FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 2946044 · Received January 7, 2013

Report

Report Number
1824206-2013-00147
Event Type
Malfunction
Date Received
January 7, 2013
Date of Event
December 21, 2012
Report Date
December 21, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO FURTHER INFO IS AVAILABLE ON THE REPAIR OF THE BED AT THIS TIME.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THAT THE BRAKES WOULD SET BUT THE BRAKE CASTERS WOULD STILL PIVOT. NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
8674 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILL-ROM, INC. 8000

Patients

Seq Age Sex Outcome Treatment
1