FDA Adverse Event Malfunction Summary report: N

AU480 CLINICAL CHEMISTRY ANALYZER WITH ISE

MDR report key: 2945873 · Received February 6, 2013

Report

Report Number
9612296-2013-00017
Event Type
Malfunction
Date Received
February 6, 2013
Date of Event
January 2, 2012
Report Date
January 10, 2013
Manufacturer
BECKMAN COULTER
Product Code
JGS
PMA / PMN Number
K981743
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) REPLACED THE PINCH VALVE, DRAIN, AND ION SELECTIVE ELECTROLYTE (ISE) TUBING SET. THE FSE ALSO REPLACED THE ELECTRODES AND SAMPLE PROBE. THE FSE PERFORMED THREE CONSECUTIVE SUCCESSFUL CALIBRATIONS AND COMPLETED MID SOLUTION REST. THE FSE REPLACED THE DETERGENT ROLLER PUMP TUBING AND DRAINED THE DETERGENT BOTTLE. THE FSE PERFORMED QUALITY CONTROL (QC) SUCCESSFULLY. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED SPORADIC LOW SODIUM (NA) PATIENT RESULTS, ON SEPARATE DAYS, INVOLVING THE AU480 CLINICAL CHEMISTRY ANALYZER WITH ISE. THE CUSTOMER STATED SUBSEQUENT TESTING RECOVERED ACCEPTABLE SODIUM RESULTS. THE CUSTOMER PERFORMED SYSTEM TROUBLESHOOTING BUT COULD NOT RESOLVE THE ISSUE. THE LOW RESULTS WERE RELEASED OUT OF THE LABORATORY AND QUESTIONED BY THE PHYSICIAN. THERE WAS NO REPORT OF PATIENT INJURY OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER INDICATED QUALITY CONTROL (QC) WAS WITHIN SPECIFICATIONS DURING THE EVENT. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
50601 AU480 CLINICAL CHEMISTRY ANALYZER WITH ISE ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JGS BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1