FDA Adverse Event Malfunction Summary report: N

COYOTE¿

MDR report key: 2945701 · Received February 6, 2013

Report

Report Number
2134265-2013-00356
Event Type
Malfunction
Date Received
February 6, 2013
Date of Event
January 8, 2013
Report Date
January 9, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K111295
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: THE COYOTE CATHETER WAS RECEIVED WITH NO ORIGINAL PACKAGING OR OTHER DEVICES. MAGNIFIED INSPECTION REVEALED NO DAMAGE. WHEN POSITIVE PRESSURE WAS APPLIED WITH AN INFLATION DEVICE FILLED WITH WATER, A STREAM OF WATER EMITTED FROM A PINHOLE 37MM FROM THE PROXIMAL EDGE OF THE DISTAL MARKERBAND. THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKER THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN AN UNSPECIFIED MODERATELY TORTUOUS VESSEL BELOW THE KNEE. THE 2.0MM X 150MM X 150CM COYOTE OTW BALLOON WAS INFLATED ONCE AND RUPTURED AT 10 ATMS. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT CONDITION IS STABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY A BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN AN UNSPECIFIED MODERATELY TORTUOUS VESSEL BELOW THE KNEE. THE 2.0MM X 150MM X 150CM COYOTE OTW BALLOON WAS INFLATED ONCE AND RUPTURED AT 10 ATMS. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT CONDITION IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
50285 COYOTE¿ CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939186201510 15593577

Patients

Seq Age Sex Outcome Treatment
1