HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2013-03025
- Event Type
- Malfunction
- Date Received
- February 6, 2013
- Date of Event
- January 28, 2013
- Report Date
- January 28, 2013
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THIS COMPLAINT FOR A REPORT OF A SYSTEM ERROR 2240 WAS CONFIRMED. THE ROOT CAUSE WAS "SUPPLY BAG DISCONNECTED WITHOUT FALLING". THE SAMPLE WAS UNAVAILABLE AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.
THE CUSTOMER CONTACTED BAXTER'S SERVICE CENTER REGARDING A SYSTEM ERROR 2240 (AIR IN TUBING) WHICH OCCURRED ON THE HOMECHOICE (HC) DURING DWELL 2. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE HOME PATIENT (HP) CYCLE THE POWER ON THE HC TO RECEIVE A SYSTEM ERROR 2367. THE TSR HAD THE HP TURN THE HC OFF AGAIN. THE TSR EXPLAINED BOTH SYSTEM ERRORS TO THE HP. THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THE PATIENT LINE HAD BEEN PROPERLY PRIMED PRIOR TO CONNECTING AND NO PATIENT EXTENSION LINES WERE IN USE. THE PATIENT LINE HAD NOT DISCONNECTED FROM THE TRANSFER SET. THE PATIENT DID NOT INITIATE THERAPY BEFORE CONNECTING AND HAD NOT DISCONNECTED PRIOR TO THE ALARM. THE SUPPLIES HAD NOT BEEN DAMAGED BY AN OUTLET PORT CLAMP OR ASSIST DEVICE. THE HP SAID THAT THE SUPPLY BAG CAME LOOSE FROM THE SUPPLY LINE AS IT WAS NOT CONNECTED PROPERLY. THE TSR ADVISED THE HP MUST START OVER WITH ALL NEW SUPPLIES ON HC. THE HP UNDERSTOOD. PROPER PROCEDURES WERE REVIEWED. THERE WERE NO SAMPLES OR LOT NUMBERS AVAILABLE. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 50472 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | HOMECHOICE |