AUTOPULSE NIMH BATTERY
Report
- Report Number
- 3003793491-2012-00555
- Event Type
- Malfunction
- Date Received
- December 31, 2012
- Date of Event
- September 3, 2012
- Report Date
- September 4, 2012
- Manufacturer
- ZOLL CIRCULATION, INC.
- Product Code
- DRM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE AUTOPULSE NIMH BATTERY (S/N (B)(4)) INVOLVED IN THE EVENT WAS RETURNED TO ZOLL CIRCULATION ON (B)(4) 2012 AND WAS ANALYZED. THE ARCHIVE DATA ANALYSIS SHOWED CUSTOMER NOT FOLLOWING PROPER BATTERY MANAGEMENT PROCEDURES OF DAILY SYSTEM CHECKS AND BATTERY SWAP. IT ALSO SHOWED THAT THIS BATTERY WAS USED REPEATEDLY ON THE INCIDENT DATE WHICH WAS NOT FRESH OUT FROM THE CHARGER THEREFORE IT HAD LOW VOLTAGE. AFTER A FULLY CHARGED AND TEST CYCLE, THIS BATTERY FAILED THE POWER OUTPUT TESTING. PROBABLE CAUSE FOR BATTERY FAILING COULD BE DUE TO AGING (MANUFACTURED IN AUGUST 2009). THE CALENDAR AGE OF THE BATTERY IS GREATER THAN 2 YEARS. THE PRODUCT LABELING INSTRUCTS THAT THE USEFUL LIFE OF EACH AUTOPULSE BATTERY IS BETWEEN 2-4 YEARS. AS WITH ALL BATTERIES, THE NIMH BATTERY CHEMISTRY HAS A FINITE CALENDAR LIFE. WITHIN THE 2-4 YEAR LIFE RANGE, THE LIFE OF EACH BATTERY IS INFLUENCED BY THE FREQUENCY OF USE AND THE FREQUENCY OF ROUTINE BATTERY MAINTENANCE. NO ADVERSE EVENT WAS REPORTED.
IT WAS REPORTED THAT THE AUTOPULSE STOPPED DURING CLINICAL USE WHEN USING BATTERY S/N (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOPULSE NIMH BATTERY | NIMH BATTERY | DRM | ZOLL CIRCULATION, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |