130 DEG AIMING ARM F/TROCHANTERIC FIXATION NAI
Report
- Report Number
- 8030965-2013-10035
- Event Type
- Injury
- Date Received
- February 1, 2013
- Date of Event
- January 21, 2011
- Report Date
- January 21, 2011
- Manufacturer
- SYNTHES GMBH
- Product Code
- HSB
- PMA / PMN Number
- EXEMPT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. A REVIEW OF THE DHR INDICATES THERE WERE NO ANOMALIES WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THIS COMPLAINT. VISUAL EVALUATION NOTED SEVERAL SIGNS OF INTENSE USE, SCRATCHES, STRESS MARKS AND DENTS ALL OVER THE INSTRUMENT. DURING FUNCTIONAL EVALUATION, THE DEVICE DID TARGET AS REQUIRED AND NO DEVIATION TO THE SPECIFICATION WAS FOUND.
DURING THE TFN PROCEDURE, DRILL BIT HIT THE FIXATION NAIL AS IT WAS NOT LINING UP CORRECTLY. SURGEON TRIED INSTRUMENTS FROM TWO DIFFERENT SETS INTERCHANGEABLE: 2 PROTECTION SLEEVES (357.386), 2 DRILL SLEEVES (357.389), 2 TROCARS (357.387), 2 OF 130 DEG AIMING ARMS (357.366) AND 2 INSERTION HANDLES (357.411). ALSO ONE OF THE HELICAL BLADE COUPLING SCREWS (357.377) WAS DEFORMED AND WOULD NOT ACCEPT THE HEXAGONAL FLEXIBLE SCREWDRIVER - SCREWDRIVER DID NOT ENGAGE THE HEXAGONAL RECESS OF THE SCREW. THE DEFORMED COUPLING SCREW WAS DISCARDED. THERE WAS NO PATIENT IMPACT; THE DOCTOR REALIZED THE TROCARS WERE GOING POSTERIOR AND GAVE THE JIG A GOOD PULL ANTERIOR. THIS ADDED ABOUT 20-30 MINUTES TO THE PROCEDURE. THIS COMPLAINT IS 2 OF 12 FOR THE SAME EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 44468 | 130 DEG AIMING ARM F/TROCHANTERIC FIXATION NAI | HSB | SYNTHES GMBH | 4747750 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |