FDA Adverse Event Injury Summary report: N

8.0MM/4.0MM DRILL SLEEVE 164MM

MDR report key: 2944983 · Received February 1, 2013

Report

Report Number
8030965-2013-10045
Event Type
Injury
Date Received
February 1, 2013
Date of Event
January 21, 2011
Report Date
January 21, 2011
Manufacturer
SYNTHES GMBH
Product Code
HTW
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. A REVIEW OF THE DHR INDICATES THERE WERE NO ANOMALIES WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THIS COMPLAINT. VISUAL EVALUATION NOTED SEVERAL SIGNS OF INTENSE USE, SCRATCHES, STRESS MARKS AND DENTS ALL OVER THE INSTRUMENT. DURING FUNCTIONAL EVALUATION THE DEVICE DID TARGET AS REQUIRED AND NO DEVIATION TO THE SPECIFICATION WAS FOUND. THE INVESTIGATION DETERMINED THAT APPLYING EXTREME FORCES ONTO THE CONSTRUCTION DURING A PROCEDURE, WHICH CAN LEAD TO A MISALIGNMENT.

Description of Event or Problem · 1

DURING THE TFN PROCEDURE, DRILL BIT HIT THE FIXATION NAIL AS IT WAS NOT LINING UP CORRECTLY. SURGEON TRIED INSTRUMENTS FROM TWO DIFFERENT SETS INTERCHANGEABLY: 2 PROTECTION SLEEVES (357.386), 2 DRILL SLEEVES (357.389), 2 TROCARS (357.387), 2 OF 130 DEG AIMING ARMS (357.366) AND 2 INSERTION HANDLES (357.411). ALSO ONE OF THE HELICAL BLADE COUPLING SCREWS (357.377) WAS DEFORMED AND WOULD NOT ACCEPT THE HEXAGONAL FLEXIBLE SCREWDRIVER- SCREWDRIVER DID NOT ENGAGE THE HEXAGONAL RECESS OF THE SCREW. THE DEFORMED COUPLING SCREW WAS DISCARDED. THERE WAS NO PATIENT IMPACT; THE DOCTOR REALIZED THE TROCARS WERE GOING POSTERIOR AND GAVE THE JIG A GOOD PULL ANTERIOR. THIS ADDED ABOUT 20-30 MINUTES TO THE PROCEDURE. THIS COMPLAINT IS 9 OF 12 FOR THE SAME EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
44846 8.0MM/4.0MM DRILL SLEEVE 164MM HTW SYNTHES GMBH 4604746

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention