FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2944769 · Received February 5, 2013

Report

Report Number
3004209178-2013-90833
Event Type
Malfunction
Date Received
February 5, 2013
Date of Event
December 27, 2012
Report Date
January 31, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE MOTOR SUPPORT DISK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN SQUIRTED OUT DURING THE MANUAL PRIME PROCESS, BUT THE CUSTOMER WAS DISCONNECTED DURING PRIME. IT WAS STATED THAT THE PISTON CONTINUES MOVING FORWARD AFTER THE RESERVOIR MAKES CONTACT AND INSULIN SQUIRTS OUT. IT WAS MENTIONED THAT THE NUMBERS DID NOT APPEAR ON THE SCREEN, AND THE BEEPS COULD NOT BE HEARD. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47821 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWL

Patients

Seq Age Sex Outcome Treatment
1