FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 2944691 · Received February 5, 2013

Report

Report Number
1061932-2013-00173
Event Type
Malfunction
Date Received
February 5, 2013
Date of Event
January 10, 2013
Report Date
January 10, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) OBSERVED THE PROBE RINSE BLOCK AND PROBE WASTE NOT DRAINING AND NOTED PINCH VALVE VL13 WAS NOT ACTUATING. THE FSE NOTED THE PRESSURE TUBING TO PINCH VALVE VL13 HAD COME OFF. THE FSE REPLACED THE TUBING AND VERIFIED PERFORMANCE. SERVICE ACTIVITY PERFORMED WAS VERIFIED TO MEET THE SPECIFIED REQUIREMENTS PER ESTABLISHED PROCEDURES. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED FLUID LEAKED INSIDE THE INSTRUMENT IN THE SECONDARY MODE AREA INVOLVING THE COULTER LH 750 HEMATOLOGY ANALYZER. THE CUSTOMER HAD ON PROPER PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, A LABORATORY COAT, AND EYE PROTECTION AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. PATIENT RESULTS WERE NOT IMPACTED. THE CUSTOMER HAS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN AT THE FACILITY. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47838 COULTER® LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients