FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CURVED SCISSORS INSTRUMENT

MDR report key: 2944620 · Received February 5, 2013

Report

Report Number
2955842-2013-00381
Event Type
Malfunction
Date Received
February 5, 2013
Date of Event
January 9, 2013
Report Date
January 9, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE BLADE ON THE INSTRUMENT HAS A CHIP. ENGINEERING CONCLUDED THAT THE DAMAGE IS LIKELY DUE TO MISHANDLING. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SET-UP OF A DA VINCI SI SURGICAL PROCEDURE, IT WAS NOTED THAT THE BLADE ON THE MONOPOLAR CURVED SCISSORS INSTRUMENT HAD A CHIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
48936 MONOPOLAR CURVED SCISSORS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-10 M11120409 462

Patients

Seq Age Sex Outcome Treatment
1 DAV SI SURG SYSTEM, MCS TIP COVER ACCESSORY & ESU