FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2944567 · Received February 5, 2013

Report

Report Number
2032227-2013-00466
Event Type
Injury
Date Received
February 5, 2013
Date of Event
January 22, 2013
Report Date
January 22, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED HAVING AN URGENT CARE VISIT DUE TO HIS HIGH BLOOD GLUCOSE OF 560MG/DL. TROUBLESHOOTING WAS PERFORMED. THE TIME, DATE, AND BOLUS WIZARD SETTINGS WERE CORRECT. REVIEWED THE ALARM HISTORY AND FOUND LOW BATTERY AND LOW RESERVOIR ALARMS. ASSISTED THE CALLER TO RUN A MANUAL PRIME AND THE INSULIN DID EXIT. ADVISED THE CUSTOMER TO CALL BACK WHEN THE TUBING CLAMP ARRIVES TO CONTINUE TESTING. INSTRUCTED THE CALLER TO REMOVE THE CANNULA AND IT WAS NOT BENT. SUGGESTED THE CUSTOMER TO CHANGE THE ENTIRE INFUSION SET AND RESERVOIR. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
48647 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAS

Patients

Seq Age Sex Outcome Treatment
1 30 YR Hospitalization