FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2944436 · Received February 5, 2013

Report

Report Number
2024168-2013-00659
Event Type
Injury
Date Received
February 5, 2013
Date of Event
January 9, 2013
Report Date
January 15, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. ESTIMATED DATE OF OCCURRENCE, EXACT DATE WAS NOT REMEMBERED BY HOSPITAL. PER THE INSTRUCTIONS FOR USE, THE SAFETY AND EFFECTIVENESS OF THE PROGLIDE DEVICE HAVE NOT BEEN ESTABLISHED FOR PATIENTS WITH FEMORAL ARTERY CALCIUM WHICH IS FLUOROSCOPICALLY VISIBLE AT THE ACCESS SITE. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE PLUNGER, LINK, AND SUTURE WERE NOT RETURNED. WITHOUT THE RETURN OF THESE COMPONENTS, THE SCOPE OF THE INVESTIGATION WAS LIMITED AND THE REPORTED SUTURE BREAK COULD NOT BE CONFIRMED. ANALYSIS OF THE RETURNED DEVICE COMPONENTS DID NOT OBSERVE A FAILURE. THE CAUSE OF THE INCIDENT COULD NOT BE DETERMINED. BASED ON VISUAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE MILDLY CALCIFIED RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE WITH A 6F SHEATH AFTER A CORONARY INTERVENTIONAL PROCEDURE. REPORTEDLY, AFTER RETRACTING THE PLUNGER A CUFF MISS OCCURRED. HEMOSTASIS WAS ACHIEVED USING MANUAL ARTERIAL COMPRESSION. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WAS REPORTED. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
48549 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 20928J2

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 6F, HEPARIN