FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2944380 · Received February 5, 2013

Report

Report Number
0001811755-2013-00185
Event Type
Malfunction
Date Received
February 5, 2013
Date of Event
January 18, 2013
Report Date
January 22, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, THE DRIVESHAFT ASSEMBLY WAS FOUND TO BE CORRODED. INCREASED FRICTION DUE TO THE CORRODED DRIVESHAFT IS A PROBABLE CAUSE OF THE REPORTED OVERHEATING.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION HAS BEEN PERFORMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY, THE DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY, THE DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
48457 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1