FDA Adverse Event
Malfunction
Summary report: N
REMB SAG SAW
MDR report key: 2944375
·
Received February 5, 2013
Report
- Report Number
- 0001811755-2013-00184
- Event Type
- Malfunction
- Date Received
- February 5, 2013
- Date of Event
- January 18, 2013
- Report Date
- January 18, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE MOTOR WAS DISCOVERED TO BE DAMAGED, WHICH IS A PROBABLE CAUSE OF THE REPORTED BIAS CURRENT ERROR.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY, THE SAW CAUSED A BIAS CURRENT WARNING TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 48926 | REMB SAG SAW | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | STRYKER INSTRUMENTS-KALAMAZOO |