Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FOUND THAT THE DIODE D7 WAS DETERMINED TO BE LEAKY AND WAS OUT OF THE SPECIFICATION. NEW PCB BOARD WAS REPLACED TO SOLVE THE ISSUE. (B)(4).

Description of Event or Problem · 1

INFORMATION RECEIVED THAT PUMP ALARMS ERROR CODE 13.