FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2944189 · Received February 5, 2013

Report

Report Number
2024168-2013-00639
Event Type
Injury
Date Received
February 5, 2013
Date of Event
January 13, 2013
Report Date
January 15, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HK
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EXPERIENCE FOR THIS COMPLAINT WAS CONFIRMED. BASED ON THE INVESTIGATION, THE CAUSE FOR THIS COMPLAINT COULD NOT BE DETERMINED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FROM THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE WITH A 6FR SHEATH, AFTER A PERCUTANEOUS CORONARY INTERVENTION (PCI) PROCEDURE. REPORTEDLY, DURING THE PROCEDURE, THE SHAFT KINKED NEAR THE GUIDE WIRE EXIT PORT. MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE ENTIRE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47664 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 21113J1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 6FRHEPARIN