STERRAD 100S STERILIZER
Report
- Report Number
- 2084725-2013-00055
- Event Type
- Malfunction
- Date Received
- February 5, 2013
- Date of Event
- January 23, 2013
- Report Date
- January 23, 2013
- Manufacturer
- ADVANCED STERILIZATION PRODUCTS
- Product Code
- MLR
- PMA / PMN Number
- K991999
- Removal / Correction Number
- Z-0744-0746-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER WAS DISPATCHED TO CUSTOMER SITE. ODOR/SMELL WAS CONFIRMED. THE CATALYTIC DECOMP FILTER AND FILTER WERE REPLACED. UNIT MEETS SPECIFICATIONS AND WAS RETURNED TO SERVICE ON (B)(4) 2013.
CORRECTION: SEX IS UNKNOWN. MANUFACTURER DATE: 11/21/2002. ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD, SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, AND SYSTEM HAZARD AND USER MISUSE ANALYSIS. THE DHR (DEVICE HISTORY RECORD) WAS REVIEWED AND INDICATED NO ANOMALIES THAT COULD HAVE CONTRIBUTED TO THE CUSTOMER'S EXPERIENCED "ODOR/SMELLS" ISSUE. THE UNIT MET SPECIFICATION AT THE TIME OF ITS RELEASE. THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST 6 MONTHS (07/27/2012 ¿ 01/23/2013) DID NOT REVEAL A SIGNIFICANT TREND. THE COMPLAINT TRENDING FOR PROBLEM CODE ¿ODOR/SMELLS¿ IDENTIFIED A SIGNIFICANT TREND OVER THE PAST 12 MONTHS (SEPTEMBER 2012 ¿ AUGUST 2013). THE HIGHEST RISK IS CONSIDERED AS LOW AS REASONABLY PRACTICAL. THE SHUMA (SYSTEM HAZARD AND USER MISUSE ANALYSIS) DEFINES THE RISK AS LOW AS REASONABLY PRACTICAL FOR EXPOSURE TO ODOR OR ODORANTS. THE CATALYTIC CONVERTER WAS RETURNED AND TESTED. THE CATALYTIC CONVERTER EXHIBITED NO MIST OR ODOR BUT FAILED ADDITIONAL FUNCTIONAL TESTING. THE OIL MIST FILTER WAS RETURNED AND TESTED. THE OIL MIST FILTER HAD NO MIST OR ODOR OBSERVED. REVIEW OF THE TRACKING AND TRENDING DATA DID NOT IDENTIFY A TREND FOR THIS STERILIZER. AS A RESULT, ROOT CAUSE OR ASSIGNABLE CAUSE ANALYSIS COULD NOT BE PERFORMED. NO FURTHER ACTION IS REQUIRED; HOWEVER, THIS ISSUE WILL CONTINUE TO BE MONITORED.
A CUSTOMER REPORTED AN EVENT OF AN ODD SMELL (ODOR) EMITTING FROM THE STERRAD 100S. PRIOR TO EXPERIENCING THE ODOR THE UNIT HAD TWO ERRORS, THE FIRST BEING OVER PRESSURE INJECTION AND THE SECOND WAS INJECTION INTERRUPTED. THE CUSTOMER TURNED OFF THE UNIT. WHEN THEY TURNED IT BACK ON THE UNIT THEN BEGAN TO EMIT THE ODOR. THERE WAS NO REPORT OF HUMAN REACTION. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(4) 2012.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 48726 | STERRAD 100S STERILIZER | STERRAD EQUIPMENT (MLR) | MLR | ADVANCED STERILIZATION PRODUCTS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |