FDA Adverse Event Malfunction Summary report: N

STERRAD 100S STERILIZER

MDR report key: 2943891 · Received February 5, 2013

Report

Report Number
2084725-2013-00053
Event Type
Malfunction
Date Received
February 5, 2013
Date of Event
January 23, 2013
Report Date
January 23, 2013
Manufacturer
ADVANCED STERILIZATION PRODUCTS
Product Code
MLR
PMA / PMN Number
K991999
Removal / Correction Number
Z-0744-0746-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER WAS DISPATCHED TO CUSTOMER SITE. ODOR/SMELL WAS CONFIRMED. A PREVENTATIVE MAINTENANCE (PM1) WAS PERFORMED. UNIT MEETS SPECIFICATIONS AND WAS RETURNED TO SERVICE ON (B)(4) 2013.

Additional Manufacturer Narrative · 1

CORRECTION: SEX IS UNKNOWN. ASP INVESTIGATION SUMMARY: THE INVESTIGATION INCLUDED A REVIEW OF THE DEVICE HISTORY RECORD, SERVICE HISTORY, TRENDING OF THE PRODUCT MALFUNCTION CODE, AND SYSTEM HAZARD AND USER MISUSE ANALYSIS. THE DHR (DEVICE HISTORY RECORD) WAS REVIEWED AND INDICATED NO ANOMALIES THAT COULD HAVE CONTRIBUTED TO THE CUSTOMER'S EXPERIENCED "ODOR/SMELLS" ISSUE. THE UNIT MET SPECIFICATION AT THE TIME OF ITS RELEASE. THE SERVICE HISTORY FOR THIS UNIT FOR THE PAST 6 MONTHS (07/27/2012 ¿ 01/232013) DID NOT REVEAL A SIGNIFICANT TREND. THE COMPLAINT TRENDING FOR PROBLEM CODE ¿ODOR/SMELLS¿ IDENTIFIED A SIGNIFICANT TREND OVER THE PAST 12 MONTHS (SEPTEMBER 2012 ¿ AUGUST 2013). THIS RISK IS CONSIDERED AS LOW AS REASONABLY PRACTICAL. THE SHUMA (SYSTEM HAZARD AND USER MISUSE ANALYSIS) DEFINES THE RISK AS LOW AS REASONABLY PRACTICAL FOR EXPOSURE TO ODOR OR ODORANTS. NO PARTS WERE RETURNED FOR FURTHER EVALUATION. REVIEW OF THE TRACKING AND TRENDING DATA DID NOT IDENTIFY A TREND FOR THIS STERILIZER. AS A RESULT, ROOT CAUSE OR ASSIGNABLE CAUSE ANALYSIS COULD NOT BE PERFORMED. NO FURTHER ACTION IS REQUIRED; HOWEVER, THIS ISSUE WILL CONTINUE TO BE MONITORED.

Description of Event or Problem · 1

A CUSTOMER REPORTED AN EVENT OF A STRONG CHEMICAL SMELL (ODOR) EMITTING FROM THE STERRAD 100S. THERE WAS NO REPORT OF HUMAN REACTION. NO LOAD WAS AFFECTED. AN ASP FIELD SERVICE ENGINEER WAS DISPATCHED TO ASSESS THE UNIT ONSITE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION REPORT SUBSEQUENT TO A SERIOUS INJURY EVENT DATED (B)(4) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47891 STERRAD 100S STERILIZER STERRAD EQUIPMENT (MLR) MLR ADVANCED STERILIZATION PRODUCTS NA

Patients

Seq Age Sex Outcome Treatment
1