FDA Adverse Event Malfunction Summary report: N

GYNECARE X-TRACT MORCELLATOR

MDR report key: 2943869 · Received February 5, 2013

Report

Report Number
2210968-2013-00761
Event Type
Malfunction
Date Received
February 5, 2013
Date of Event
December 27, 2012
Report Date
January 16, 2013
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K993801
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. UPON EVALUATION, THERE WAS A MISALIGNMENT OF COUPLER AND CPC CONNECTOR.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2012. DURING THE PROCEDURE, IT WAS DIFFICULT TO CONNECT THE HANDPIECE TO THE DEVICE, AND WHEN THEY DID CONNECT THE HANDPIECE, IT KEPT FALLING OUT SO THEY HAD TO HOLD THE HANDPIECE CONNECTOR IN THE DEVICE CONNECTION SITE. IT WAS ALSO REPORTED THAT THE DEVICE WAS NOT PROVIDING ENOUGH POWER TO THE HANDPIECE. THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
48586 GYNECARE X-TRACT MORCELLATOR LAPAROSCOPIC MORCELLATOR HET ETHICON, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention