GYNECARE X-TRACT MORCELLATOR
Report
- Report Number
- 2210968-2013-00761
- Event Type
- Malfunction
- Date Received
- February 5, 2013
- Date of Event
- December 27, 2012
- Report Date
- January 16, 2013
- Manufacturer
- ETHICON, INC.
- Product Code
- HET
- PMA / PMN Number
- K993801
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.
(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. UPON EVALUATION, THERE WAS A MISALIGNMENT OF COUPLER AND CPC CONNECTOR.
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2012. DURING THE PROCEDURE, IT WAS DIFFICULT TO CONNECT THE HANDPIECE TO THE DEVICE, AND WHEN THEY DID CONNECT THE HANDPIECE, IT KEPT FALLING OUT SO THEY HAD TO HOLD THE HANDPIECE CONNECTOR IN THE DEVICE CONNECTION SITE. IT WAS ALSO REPORTED THAT THE DEVICE WAS NOT PROVIDING ENOUGH POWER TO THE HANDPIECE. THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 48586 | GYNECARE X-TRACT MORCELLATOR | LAPAROSCOPIC MORCELLATOR | HET | ETHICON, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |