FDA Adverse Event Injury Summary report: N

KX+ DURAT LS INSERT MED 8MM

MDR report key: 2943862 · Received February 5, 2013

Report

Report Number
0002249697-2013-00566
Event Type
Injury
Date Received
February 5, 2013
Date of Event
January 16, 2013
Report Date
January 23, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
JWH
PMA / PMN Number
K921640
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE DEVICE HISTORY REVIEW CONCLUDED THE DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THE COMPLAINT HISTORY REVIEW CONCLUDED THERE HAVE BEEN NO OTHER EVENTS FOR THE LOT REFERENCED. VISUAL INSPECTION WAS PERFORMED AS PART OF THE MATERIAL ANALYSIS. THE REPORT CONCLUDED: THE POSTERIOR LOCKING TABS OF THE TIBIAL BASEPLATE SHOWED SIGNS OF CONTACT WITH THE FEMORAL COMPONENT. IT IS PROBABLE THAT THE MAJOR PART OF THE DAMAGE TO THE POSTERIOR LOCKING TABS MUST HAVE OCCURRED PRIOR TO THE REVISION SURGERY OF 2012 BECAUSE THERE WAS NO POSTERIOR DAMAGE TO THE RETURNED TIBIAL INSERT. THE ARTICULATING SURFACE OF THE FEMORAL COMPONENT SHOWED SIGNS OF SCRATCHES AND GOUGES FROM CONTACT WITH THE POSTERIOR LOCKING TABS OF THE TIBIAL BASEPLATE AND FROM CONTACT WITH THE EMBEDDED METAL FRAGMENT IN THE TIBIAL INSERT. THE POSTERIOR LOCKING TAB OF THE TIBIAL BASEPLATE BROKE IN FATIGUE DUE TO REPETITIVE LOADING BY THE FEMORAL COMPONENT. FINAL FRACTURE OF THE FRAGMENT OF THE LOCKING TAB MAY HAVE OCCURRED PRE OR POST THE REVISION SURGERY OF 2012. THE BROKEN FRAGMENT OF THE POSTERIOR LOCKING TAB ENDED UP BEING EMBEDDED IN THE ARTICULATING SURFACE OF THE TIBIAL INSERT. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON ANY OF THE COMPONENTS. MEDICAL RECORDS RECEIVED AND EVALUATION: SEVERE POLY WEAR PRESENT PRIOR TO REVISION IN 2012 WITH FLEXION INSTABILITY LEADING TO METAL-METAL CONTACT BETWEEN FEMORAL COMPONENT AND THE POSTERIOR RIM OF THE BASEPLATE ONCE ENOUGH POLY HAD BEEN WORN OFF. FATIGUE FRACTURE OF THE POSTERIOR LOCKING TAB MOST PROBABLY TOOK PLACE BEFORE THE REVISION PROCEDURE IN 2012 WITH THE BROKEN PIECE SHIELDED FROM VIEW BOTH DURING REVISION SURGERY AND ON X-RAY DUE TO THE METAL SHIELDING OF THE LARGE COMPONENTS AROUND. AS SUCH, ROOT CAUSE OF FAILURE IN THIS PI CASE IS PROCEDURE-RELATED AND NOT DEVICE-RELATED WITH NO FURTHER PATIENT-RELATED FACTORS EVIDENT. THE EVENT WAS CONFIRMED. THE INVESTIGATION CONCLUDED THAT THE METAL FRAGMENT FOUND EMBEDDED IN THE BEARING SURFACE OF THE TIBIAL INSERT WAS CAUSED BY THE NON-DETECTION OF THE DAMAGE TO THE POSTERIOR ASPECT OF THE BASEPLATE DURING THE PREVIOUS REVISION SURGERY IN (B)(6) 2012. AS SUCH THE ROOT CAUSE IS DETERMINED TO BE USER RELATED IN THAT THE DAMAGE TO THE POSTERIOR ASPECT OF THE BASEPLATE WAS NOT DETECTED SOONER. THE EVENT IS NOT DEVICE-RELATED.

Description of Event or Problem · 1

PATIENT PRESENT TO SURGEONS OFFICE WITH SWOLLEN KNEE POST REPLACEMENT OF INSERTS TO FIND A PIECE OF METAL SHOWING ON XRAY. SURGERY TO EXPLORE WAS SET.

Description of Event or Problem · 1

PATIENT PRESENT TO SURGEONS OFFICE WITH SWOLLEN KNEE POST REPLACEMENT OF INSERTS TO FIND A PIECE OF METAL SHOWING ON XRAY. SURGERY TO EXPLORE WAS SET.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
48584 KX+ DURAT LS INSERT MED 8MM IMPLANT JWH STRYKER ORTHOPAEDICS-MAHWAH LBG927

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention