FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2943829 · Received February 5, 2013

Report

Report Number
1416980-2013-02958
Event Type
Malfunction
Date Received
February 5, 2013
Date of Event
October 29, 2012
Report Date
October 29, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE ROOT CAUSE OF THE POWER SUPPLY FAILURE WAS DETERMINED TO BE DAMAGE TO POWER BOARD. TO CORRECT THE CONDITION, THE POWER BOARD WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A POWER SUPPLY FAILURE. "THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY". THIS WAS NOT REPORTED BY THE CUSTOMER. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
49002 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1