FLOGARD
Report
- Report Number
- 1416980-2013-02945
- Event Type
- Malfunction
- Date Received
- February 5, 2013
- Date of Event
- November 1, 2012
- Report Date
- November 14, 2012
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. THE ROOT CAUSE OF THE DOWNSTREAM OCCLUSION FALSE ALARM WAS DETERMINED TO BE DAMAGE TO THE DOOR LATCH ROLLER. TO CORRECT THE CONDITION THE DOOR LATCH ROLLER WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP MDR WILL BE SENT.
DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DOWNSTREAM OCCLUSION FALSE ALARM. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS WAS NOT REPORTED BY THE CUSTOMER. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.