FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX¿

MDR report key: 2943617 · Received February 5, 2013

Report

Report Number
2134265-2013-00259
Event Type
Malfunction
Date Received
February 5, 2013
Date of Event
January 4, 2013
Report Date
January 7, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MFR.: MAGNIFIED INSPECTION REVEALED NO DAMAGE. WHEN POSITIVE PRESSURE WAS APPLIED WITH AN INFLATION DEVICE FILLED WITH WATER, A STREAM OF WATER EMITTED FROM A PINHOLE IN THE BALLOON WALL 5MM FROM THE PROXIMAL EDGE OF THE DISTAL MARKERBAND. THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKER THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION A BALLOON RUPTURE OCCURRED. THE LESION BEING TREATED WAS AN AREA OF INSTENT RESTENOSIS OF AN UNKNOWN STENT LOCATED IN THE MODERATELY TORTUOUS LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE PHYSICIAN DILATED THE LESION WITH A 2.5 X 20MM NON-BSC BALLOON CATHETER THEN ADVANCED A 15MM X 2.75MM NC QUANTUM APEX BALLOON CATHETER TO THE TARGET LESION. UPON THE FIRST INFLATION AT 20ATMS THE BALLOON RUPTURED. THE DEVICE WAS SUCCESSFULLY REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
47854 NC QUANTUM APEX¿ CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912415270 15160828

Patients

Seq Age Sex Outcome Treatment
1 2.5 X 20MM FUKUDA BALLOON CATHETER