FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 2943500 · Received January 9, 2013

Report

Report Number
1824206-2013-00224
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
January 6, 2013
Report Date
January 6, 2013
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE CARDIOPULMONARY RESUSCITATION DAMPENER TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECHNICIAN ALLEGED THE HEAD OF BED WOULD SLAM DOWN QUICKLY WITHOUT RESISTANCE. NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12986 ADVANTA BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment
1