FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 2943475 · Received January 8, 2013

Report

Report Number
1314492-2013-00014
Event Type
Malfunction
Date Received
January 8, 2013
Report Date
December 12, 2012
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Removal / Correction Number
1314492-080712-002-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE BAXTER DEVICE EVAL FOUND THE #2, #3, #4, #5, #6, #7, #8, #9 AND (.) KEYS TO BE INOPERABLE. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #2 KEY WILL OCCUR FOLLOWING THE SELECTED KEY'S FUNCTION (EG, WHEN THE #1 KEY IS PRESSED 12 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE ABC KEY IS SELECTED, AD WILL BE DISPLAYED INTERFERING WITH THE (B)(4)). FURTHER EVAL REVEALED VISIBLE KEYPAD DAMAGE AT THE #2, #5, AND #8 KEYS CAUSED BY AN EXTERNAL FORCE. BAXTER'S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. AT THIS TIME, THE DATE OF EVENT IS UNK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD ON A PUMP WAS INOPERABLE. THE CUSTOMER STATED THAT THE "KEYPAD DOUBLE TAPS EVERY TIME A KEY IS PRESSED (IE, PUSHING "7"), "77" APPEARS."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9612 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1