XIA 3 TITANIUM BLOCKER
Report
- Report Number
- 3005525032-2013-00001
- Event Type
- Malfunction
- Date Received
- February 5, 2013
- Date of Event
- January 7, 2013
- Report Date
- January 8, 2013
- Manufacturer
- STRYKER SPINE-SWITZERLAND
- Product Code
- KWP
- PMA / PMN Number
- K071373
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF MADE AVAILABLE WILL BE REPORTED IN A SUPPLEMENTAL. METHOD, RESULT, AND CONCLUSION CODES WILL BE MADE AVAILABLE FOLLOWING AN ENGINEERING EVALUATION.
RESULTS: VISUAL INSPECTION: THREE BLOCKERS WERE RETURNED IN TWO SEPARATE PACKAGES. BLOCKER FOUND IN PACKAGING MARKED "S1 RIGHT BLOCKER": BOTTOM SURFACE IS DEFORMED. THE MARKS LEFT BY A ROD ARE VISIBLE. THEY ARE NOT VERY WELL DEFINED. ONE DEFORMATION IS DEEPER THAT THE OPPOSITE ONE. THIS CAN SUGGEST ABOUT INCORRECTLY PLACED ROD IN THE SCREW TULIP BEFORE TIGHTENING. IN ADDITION, THE SHAPE OF THE OBSERVED DEFORMATIONS IS COMMON FOR THE SITUATION WHEN ROD IS NOT SUFFICIENTLY ENGAGED INTO SCREW TULIP. THE FIRST THREAD IS DEFORMED. SUCH DEFORMATION COULD HAVE OCCURRED AS A RESULT OF MISALIGNMENT OF BLOCKER WITH SCREW TULIP DURING ITS INSERTION. TWO BLOCKERS FOUND IN PACKAGING MARKED "CROSS THREADING DAMAGED BLOCKERS": FIRST THREADS OF BOTH BLOCKERS ARE DAMAGED. HOWEVER, ON BOTH BLOCKERS ONE CAN OBSERVE THE IMPRINTS OF A ROD ON THE BOTTOM SURFACES. IT MEANS THAT DESPITE THE CROSS THREADING, BOTH OF THEM WERE DRIVEN INTO THE SCREW TULIP AND TRIED TO BE TIGHTENED. THE WALLS OF HEXAGONAL INTERFACE OF BOTH BLOCKERS ARE NOT DAMAGED. DEVICE HISTORY REVIEW: NO NON CONFORMITIES OR DEVIATIONS LINKED WITH REPORTED EVENT WERE NOTED DURING MANUFACTURING PROCESS. COMPLAINT HISTORY: IN TOTAL BLOCKER CROSS THREADING OF BLOCKER DEFORMATION ISSUE IS CONFIRMED IN 16 OF THE 24 COMPLAINTS IN A TOTAL OF 25 UNITS. RISK ANALYSIS: THE POTENTIAL EFFECT "BLOCKER FAILS OR TULIP/SCREW BECOMES COMPROMISED OR TULIP SEPARATES FROM SCREW OR TULIP COULD BECOME SPLAYED" OF FAILURE MODE "DOES NOT LOOK AT ARROWS DURING FINAL TIGHTENING" AT FINAL TIGHTENING STEP OF SURGERY IS TAKEN INTO ACCOUNT IN CURRENT A FMEA WITH OCCURRENCE 7 AND CORRESPONDING SEVERITY 7/S2. CONCLUSION: THREE BLOCKERS WERE RECEIVED FOR INVESTIGATION. AT VISUAL INSPECTION DAMAGE OF FIRST THREAD FROM BOTTOM SIDE (CROSS THREADING) WAS CONFIRMED FOR ALL BLOCKERS. IN ADDITION, THE DEFORMATION OF BOTTOM SIDE (RESULT OF FINAL TIGHTENING) WAS ALSO OBSERVED ON ALL THREE BLOCKERS. HOWEVER, ON ALL THREE BLOCKERS THIS DEFORMATION IS NOT WELL DEFINED. IT CAN SIGNIFY THAT THE TIGHTENING WAS NOT PERFORMED UNTIL 12NM OR THAT THE ROD WAS NOT WELL INSTALLED INTO TULIP OF CORRESPONDING SCREW OR WAS ACUTE BENT. THE CROSS THREADING TYPICALLY RESULTS FROM MISALIGNMENT OF BLOCKER DURING ITS INSERTION AND TIGHTENING. IN ADDITION, ONE OF THE FOLLOWING ROD POSITIONS IN THE SCREW TULIP CAN ALSO LEAD TO THE CROSS THREADING AND BLOCKER DAMAGE: THE ROD IS NOT HORIZONTALLY PLACED INTO THE SCREW HEAD. THE ROD IS HIGH IN THE SCREW HEAD. AN ACUTE CONVEX OR CONCAVE BEND IS CONTOURED INTO THE ROD.
IT WAS REPORTED THAT, "DURING XIA3 SURGERY, 2 BLOCKER WORN. WHEN THE FINAL TIGHTENING OF LEFT S2AI SCREW, THE SURGEON HEARD A STRANGE SOUND. BECAUSE IT SEEMED CROSSING THREAD, HE CHANGED TO OTHER BLOCKER AND DID THE FINAL TIGHTENING. THEN WHEN THE FINAL TIGHTENING OF RIGHT S1 SCREW, THE SURGEON HEARD A STRANGE SOUND AGAIN. BECAUSE THE BLOCKER CONTINUED WEARING, HE CHANGED TO OTHER SCREW AND BLOCKER. FINALLY HE COULD DO THE FINAL TIGHTENING OF ALL SCREWS."
IT WAS REPORTED THAT, "DURING XIA3 SURGERY, 2 BLOCKER WORN. WHEN THE FINAL TIGHTENING OF LEFT S2AI SCREW, THE SURGEON HEARD A STRANGE SOUND. BECAUSE IT SEEMED CROSSING THREAD, HE CHANGED TO OTHER BLOCKER AND DID THE FINAL TIGHTENING. THEN WHEN THE FINAL TIGHTENING OF RIGHT S1 SCREW, THE SURGEON HEARD A STRANGE SOUND AGAIN. BECAUSE THE BLOCKER CONTINUED WEARING, HE CHANGED TO OTHER SCREW AND BLOCKER. FINALLY HE COULD DO THE FINAL TIGHTENING OF ALL SCREWS."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 48851 | XIA 3 TITANIUM BLOCKER | IMPLANT | KWP | STRYKER SPINE-SWITZERLAND | M3S |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | XIA 3 TITANIUM POLYAXIAL SCREW DIA 7.5 X 35 MM |