ACCU-CHEK SPIRIT
Report
- Report Number
- 2183996-2013-00048
- Event Type
- Malfunction
- Date Received
- January 21, 2013
- Date of Event
- January 9, 2013
- Report Date
- March 28, 2013
- Manufacturer
- ROCHE HEALTH SOLUTIONS INC.
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RO
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE COMPLAINT CANNOT BE VERIFIED DUE TO CARELESS HANDLING OF THE PRODUCT THE SOFT COMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY. THEREFORE MOISTURE MAY ENTER THE INSULIN PUMP AND DESTROY THE PUMP ELECTRONICS. THE BUTTONS FUNCTION WERE TESTED SUCCESSFUL AND MEET THE SPECIFICATION. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO CARELESS HANDLING OF THE PRODUCT.
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
ON (B)(4) 2013, IT WAS REPORTED THAT THE UP AND DOWN BUTTONS ON THE PT'S INFUSION DEVICE WERE NOT FUNCTIONING. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WAS REPLACED AND REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 28929 | ACCU-CHEK SPIRIT | LZG | ROCHE HEALTH SOLUTIONS INC. | 00700006862 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INSULIN| INSULIN INFUSION PUMP| RELATED ACCESSORIES |