ACCU-CHEK SPIRIT
Report
- Report Number
- 2183996-2012-02129
- Event Type
- Malfunction
- Date Received
- January 18, 2013
- Date of Event
- October 26, 2012
- Report Date
- March 26, 2013
- Manufacturer
- ROCHE HEALTH SOLUTIONS INC.
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- Z-1646-2009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HU
- Reporter Occupation
- NOT APPLICABLE
Narratives
PRODUCT WAS RECEIVED ON (B)(4) 2012. THE SOFTCOMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY AND RESTRICTED HANDLING OF THE PUMP IS A POSSIBLE IMPLICATION. THEREFORE, MOISTURE ENTERED THE INSULIN PUMP AND DESTROYED THE FUNCTIONALITY OF THE BUTTONS AND THE PUMP ELECTRONICS. THE UP BUTTON DOES NOT RESPOND TO COMMANDS DUE TO AN EXTERNAL MECHANIC DAMAGE.
THIS INCIDENT OCCURRED OUTSIDE OF THE UNITED STATES. INFO CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFO IS OBTAINED IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
THE PT REPORTED AN ISSUE WITH THE UP/DOWN BUTTON OF THE INFUSION DEVICE. NO SERIOUS INJURY WAS REPORTED. THE INFUSION DEVICE WAS REQUESTED TO BE RETURNED FOR EVALUATION. EVALUATION OF THE DEVICE FOUND THAT THE UP/DOWN BUTTON WAS DEFECTIVE. DURING A TELECONFERENCE ((B)(4) 2012) WITH PERSONNEL FROM THE OSB, A REQUEST WAS MADE TO SUBMIT MEDWATCH REPORTS FOR ALL OUS COMPLAINTS RELATED TO THE ACCU-CHECK SPIRIT RECALL (Z-1646-2009) RECEIVED AFTER CLOSURE OF THE RECALL ON 07/03/2012.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 27758 | ACCU-CHEK SPIRIT | LZG | ROCHE HEALTH SOLUTIONS INC. | 00700006862 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | INSULIN| RELATED ACCESSORIES| INSULIN INFUSTION PUMP |