FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK SPIRIT

MDR report key: 2943064 · Received January 18, 2013

Report

Report Number
2183996-2012-02129
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
October 26, 2012
Report Date
March 26, 2013
Manufacturer
ROCHE HEALTH SOLUTIONS INC.
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
Z-1646-2009
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HU
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT WAS RECEIVED ON (B)(4) 2012. THE SOFTCOMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY AND RESTRICTED HANDLING OF THE PUMP IS A POSSIBLE IMPLICATION. THEREFORE, MOISTURE ENTERED THE INSULIN PUMP AND DESTROYED THE FUNCTIONALITY OF THE BUTTONS AND THE PUMP ELECTRONICS. THE UP BUTTON DOES NOT RESPOND TO COMMANDS DUE TO AN EXTERNAL MECHANIC DAMAGE.

Additional Manufacturer Narrative · 1

THIS INCIDENT OCCURRED OUTSIDE OF THE UNITED STATES. INFO CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFO IS OBTAINED IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

THE PT REPORTED AN ISSUE WITH THE UP/DOWN BUTTON OF THE INFUSION DEVICE. NO SERIOUS INJURY WAS REPORTED. THE INFUSION DEVICE WAS REQUESTED TO BE RETURNED FOR EVALUATION. EVALUATION OF THE DEVICE FOUND THAT THE UP/DOWN BUTTON WAS DEFECTIVE. DURING A TELECONFERENCE ((B)(4) 2012) WITH PERSONNEL FROM THE OSB, A REQUEST WAS MADE TO SUBMIT MEDWATCH REPORTS FOR ALL OUS COMPLAINTS RELATED TO THE ACCU-CHECK SPIRIT RECALL (Z-1646-2009) RECEIVED AFTER CLOSURE OF THE RECALL ON 07/03/2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27758 ACCU-CHEK SPIRIT LZG ROCHE HEALTH SOLUTIONS INC. 00700006862 NA

Patients

Seq Age Sex Outcome Treatment
1 INSULIN| RELATED ACCESSORIES| INSULIN INFUSTION PUMP