ADVIA CENTAUR
Report
- Report Number
- 2432235-2012-00272
- Event Type
- Malfunction
- Date Received
- February 4, 2013
- Date of Event
- September 13, 2011
- Report Date
- August 8, 2012
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- JJE
- PMA / PMN Number
- K971418
- Removal / Correction Number
- 2432235-08/16/2012-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A SIEMENS HEALTHCARE DIAGNOSTICS FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED TO THE CUSTOMER SITE FOR INSTRUMENT EVALUATION. THE FSE CHECKED FOR BLEACH CONTAMINATION AT EACH FLUID LINE AND CONFIRMED BLEACH RESIDUAL AT THE RESUSPENSION PORT. THE FSE TESTED THE 3 WAY VALVES AND REPLACED THE WASH MANIFOLD, WASH TUBING, 3-WAY AND 2-WAY VALVES. SIEMENS INFORMED THE CUSTOMER TO PRIME THE RESUSPENSION PORT BEFORE RUNNING QC. A FIELD CORRECTION WAS IMPLEMENTED. NO FURTHER EVALUATION OF THE DEVICE IS NECESSARY. THE SYSTEM IS PERFORMING WITHIN SPECIFICATIONS. NOTE: (B)(4). SIEMENS DID NOT SEE THIS MDR POSTED ON THE FDA WEBSITE (MAUDE) AND IS RESENDING THIS MDR.
AFTER A DAILY CLEANING PROCEDURE (DCP), THE CUSTOMER STATED CA 19-9A QC RESULTS WERE LOW. THE CUSTOMER HAS BEEN PRIMING EACH FLUID LINE PRIOR TO RUNNING QC AND SAMPLES. BASED ON THE INFORMATION PROVIDED THERE WERE NO INCORRECT PATIENT RESULTS ASSOCIATED WITH THIS COMPLAINT. SIEMENS PROVIDED A TEMPORARY WORKAROUND TO THE CUSTOMER AT THE TIME OF THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 46028 | ADVIA CENTAUR | IMMUNOASSAY ANALYZER | JJE | SIEMENS HEALTHCARE DIAGNOSTICS, INC. | ADVIA CENTAUR | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |