FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2942445
·
Received February 4, 2013
Report
- Report Number
- 2531779-2013-01472
- Event Type
- Malfunction
- Date Received
- February 4, 2013
- Report Date
- January 9, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013, WITH THE FOLLOWING FINDINGS: THERE WAS NO PUMP KEYPAD DAMAGE. THE CONTRAST BUTTON WAS FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES; THE CONTRAST BUTTON HAS NO EFFECT ON THE PUMP'S INSULIN DELIVERY FUNCTION. ALL OTHER KEYPAD BUTTONS WERE TESTED AND WERE FOUND TO BE RESPONDING NORMALLY. THE KEYPAD WAS REMOVED FROM THE PUMP AND CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. UNRELATED TO THE COMPLAINT, MOISTURE WAS FOUND IN THE BATTERY COMPARTMENT. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED TO ANIMAS FOR WORN BUTTON SYMBOLS. DURING INVESTIGATION, CONTAMINATION WAS FOUND UNDER THE BUTTON CONTACTS. THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 46560 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |