FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2942445 · Received February 4, 2013

Report

Report Number
2531779-2013-01472
Event Type
Malfunction
Date Received
February 4, 2013
Report Date
January 9, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013, WITH THE FOLLOWING FINDINGS: THERE WAS NO PUMP KEYPAD DAMAGE. THE CONTRAST BUTTON WAS FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES; THE CONTRAST BUTTON HAS NO EFFECT ON THE PUMP'S INSULIN DELIVERY FUNCTION. ALL OTHER KEYPAD BUTTONS WERE TESTED AND WERE FOUND TO BE RESPONDING NORMALLY. THE KEYPAD WAS REMOVED FROM THE PUMP AND CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. UNRELATED TO THE COMPLAINT, MOISTURE WAS FOUND IN THE BATTERY COMPARTMENT. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED TO ANIMAS FOR WORN BUTTON SYMBOLS. DURING INVESTIGATION, CONTAMINATION WAS FOUND UNDER THE BUTTON CONTACTS. THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
46560 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 14 YR