LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 3005075853-2013-00463
- Event Type
- Malfunction
- Date Received
- February 4, 2013
- Date of Event
- December 18, 2012
- Report Date
- January 17, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION WITH THE JAWS PROPERLY ALIGNED. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, RETAINED AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON? I DON'T KNOW. ITS WAS DURING A LAP CHOLECYSTECTOMY. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? NO INFORMATION. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? NO INFORMATION. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? NO INFORMATION. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? NO INFORMATION. WERE ANY UNEXPECTED NOISES HEARD? ASKU. IF SO, WHEN? NO INFORMATION. DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? NO INFORMATION. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? NO. WHAT WERE THE INDICATIONS FOR SURGERY? WHAT WAS FOUND? IT DON'T KNOW ABOUT THE INDICATION OF THE CHOLECYSTECTOMY. DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? IF SO, WHAT? NO INFORMATION. DID SOMEBODY OTHER THAN THE PRIMARY SURGEON FIRE THE INSTRUMENT? IF SO, WHOM? NO INFORMATION.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CLIPS COULD NOT BE CLOSED PARTIALLY. NO FURTHER INFORMATION. NO CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 46007 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | J4CD6R |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |