FDA Adverse Event Malfunction Summary report: N

ENDOPATH** XCEL* BLADELESS TROCAR

MDR report key: 2942369 · Received February 4, 2013

Report

Report Number
3005075853-2013-00460
Event Type
Malfunction
Date Received
February 4, 2013
Date of Event
January 16, 2013
Report Date
January 17, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE SEALS OF THE UNIVERSAL SEAL ASSEMBLY TORN. A POTENTIAL CAUSE OF THIS FAILURE MAY BE THAT THE SEALS GOT DAMAGED DURING THE INSERTION/REMOVAL OF THE SHARP SURGICAL DEVICE. AS THE OBTURATOR WAS NOT RECEIVED AND IT IS THE PART THAT HAS THE BATCH STAMPED, WE DID NOT RECEIVE A BATCH OR LOT NUMBER FOR THE PRODUCT INVOLVED IN THIS COMPLAINT. THEREFORE, WE WERE UNABLE TO CHECK MANUFACTURING RECORDS FOR ANY RELATED NON-CONFORMANCES.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC DISTAL PANCREATECTOMY PROCEDURE, THE DUCK-BILLED VALVE WAS DAMAGED AND IT FELL INTO THE PATIENT. THE FRAGMENTS WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
45889 ENDOPATH** XCEL* BLADELESS TROCAR LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1