PROXIMATE** LINEAR CUTTER RELOAD - STANDARD
Report
- Report Number
- 3005075853-2013-00453
- Event Type
- Malfunction
- Date Received
- February 4, 2013
- Date of Event
- January 6, 2013
- Report Date
- January 8, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). ATTEMPTS HAVE BEEN MADE TO HAVE DEVICE RETURNED. DEVICE LOCATION IS UNKNOWN. THE DEVICE WAS NOT RECEIVED FOR ANALYSIS; THEREFORE, THE COMPLAINT COULD NOT BE CONFIRMED. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED IN THE INSTRUMENT. THE CARTRIDGE RELOAD HAD THE SWING TAB IN THE LOCKED POSITION, AND THE PROXIMAL 6 DRIVERS UP WITHOUT STAPLES; THE REMAINING DRIVERS WERE DOWN WITH STAPLES PRESENT. IT COULD NOT BE DETERMINED IF THE REPORTED INCIDENT WAS THE RESULT OF A PREMATURE LOCKOUT, OR THE RESULT OF AN INTERRUPTED FIRING STROKE. THE SWING TAB WAS RESET AND THEN THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY WITH THE RETURNED CARTRIDGE RELOAD AND IT PERFORMED AS INTENDED. WHILE NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT, IT SHOULD BE NOTED THAT 100% INSPECTION IS PERFORMED BY MEANS OF AUTOMATED VISION SYSTEM TO INSURE THE SWING TAB IS PRESENT AND UNLOCKED. IN ADDITION, PLEASE NOTE THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.
(B)(4).
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE WAS DIFFICULT TO BE FIRED AND SOME STAPLES DIDN'T COME OUT AFTER THE FIRING. ANOTHER DEVICE WAS PULLED TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 46194 | PROXIMATE** LINEAR CUTTER RELOAD - STANDARD | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | J5GZ06 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |