FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2942064
·
Received December 27, 2012
Report
- Report Number
- 1811755-2012-04353
- Event Type
- Malfunction
- Date Received
- December 27, 2012
- Date of Event
- November 19, 2012
- Report Date
- December 4, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE DEVICE EVAL, THE ROTOR WAS FOUND TO BE STICKING IN THE DRIVESHAFT, AND SEVERAL WORN AND DAMAGED COMPONENTS WERE DISCOVERED, INCLUDING BROKEN DRIVESHAFT BEARINGS. THE FRICTION CAUSED BY WORN OR DAMAGED PARTS IS A PROBABLE CAUSE OF OVERHEATING.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING AN EXTRACTION PROCEDURE AT THE USER FACILITY, THE DRILL WAS OVERHEATING. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |