CORE IMPACTION DRILL

MDR report key: 2942064 · Received December 27, 2012

Report

Report Number
1811755-2012-04353
Event Type
Malfunction
Date Received
December 27, 2012
Date of Event
November 19, 2012
Report Date
December 4, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVAL, THE ROTOR WAS FOUND TO BE STICKING IN THE DRIVESHAFT, AND SEVERAL WORN AND DAMAGED COMPONENTS WERE DISCOVERED, INCLUDING BROKEN DRIVESHAFT BEARINGS. THE FRICTION CAUSED BY WORN OR DAMAGED PARTS IS A PROBABLE CAUSE OF OVERHEATING.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN EXTRACTION PROCEDURE AT THE USER FACILITY, THE DRILL WAS OVERHEATING. BACK-UP EQUIPMENT WAS USED TO COMPLETE THE PROCEDURE. NO DELAY, NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION WERE REPORTED.