FDA Adverse Event Malfunction Summary report: N

TPS MICRO DRIVER

MDR report key: 2942063 · Received December 27, 2012

Report

Report Number
1811755-2012-04359
Event Type
Malfunction
Date Received
December 27, 2012
Date of Event
December 5, 2012
Report Date
December 5, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE DEVICE EVALUATION, THE PRINTED CIRCUIT BOARD (FLEX ASSEMBLY) WAS FOUND TO BE DAMAGED, CORROSION WAS FOUND ON THE PRINTED CIRCUIT BOARD WIRES AND THE SOCKETS, WHICH ARE PROBABLE CAUSES FOR THE REPORTED RUNNING WITHOUT USER ACTIVATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY, THE DEVICE WAS RUNNING WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TPS MICRO DRIVER ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK