FDA Adverse Event
Malfunction
Summary report: N
TPS MICRO DRIVER
MDR report key: 2942063
·
Received December 27, 2012
Report
- Report Number
- 1811755-2012-04359
- Event Type
- Malfunction
- Date Received
- December 27, 2012
- Date of Event
- December 5, 2012
- Report Date
- December 5, 2012
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE DEVICE EVALUATION, THE PRINTED CIRCUIT BOARD (FLEX ASSEMBLY) WAS FOUND TO BE DAMAGED, CORROSION WAS FOUND ON THE PRINTED CIRCUIT BOARD WIRES AND THE SOCKETS, WHICH ARE PROBABLE CAUSES FOR THE REPORTED RUNNING WITHOUT USER ACTIVATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER FACILITY, THE DEVICE WAS RUNNING WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TPS MICRO DRIVER | ERL | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |