SYSTEM 6 SAGITTAL SAW
Report
- Report Number
- 0001811755-2013-00169
- Event Type
- Malfunction
- Date Received
- February 4, 2013
- Date of Event
- January 24, 2013
- Report Date
- January 24, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
IT WAS NOTED DURING FAILURE ANALYSIS THAT THE HANDPIECE HAD HIGH CURRENT DRAW, BLADE WHIP, AND REDUCED CUTTING PERFORMANCE DUE TO A WORN ECCENTRIC BEARING. BASED ON A REVIEW OF COMPLAINT TRENDING, A WORN ECCENTRIC BEARING CAN CAUSE ALL THESE SYMPTOMS.
FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.
THE SYSTEM 6 SAGITTAL SAW WAS SENT FOR EVALUATION DUE TO HAVING BLADE WHIP WHICH CAUSED THE BLADE TO POP OUT THE INCISION DURING THE CUT TEST CONDUCTED BY A ROUTINE FIELD SERVICE MAINTENANCE VISIT. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.
THE SYSTEM 6 SAGITTAL SAW WAS SENT FOR EVALUATION DUE TO HAVING BLADE WHIP WHICH CAUSED THE BLADE TO POP OUT THE INCISION DURING THE CUT TEST CONDUCTED BY A ROUTINE FIELD SERVICE MAINTENANCE VISIT. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 46268 | SYSTEM 6 SAGITTAL SAW | INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | KIJ | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |