FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2941864 · Received February 4, 2013

Report

Report Number
0001811755-2013-00167
Event Type
Malfunction
Date Received
February 4, 2013
Date of Event
January 16, 2013
Report Date
January 17, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ALLEGED FAILURE OF OVERHEATING COULD NOT BE DUPLICATED DURING THE DEVICE EVALUATION, DUE TO A TEMPERATURE WARNING ON THE CONSOLE WHICH PREVENTED THE HANDPIECE FROM RUNNING. HOWEVER, DAMAGED BEARINGS WERE DISCOVERED IN THE DEVICE. THE WORN BEARINGS CAN CAUSE OVERHEATING.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS AT THIS TIME; IT IS NOT POSSIBLE TO DETERMINE THE CAUSE OF THE REPORTED MALFUNCTION WITHOUT AN EVALUATION OF THE DEVICE. ADDITIONAL INFORMATION WILL BE SUBMITTED WHEN THE DEVICE IS RECEIVED, AND IF ADDITIONAL INFORMATION IS RECEIVED AND REQUIRES REPORTING.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE EXTRACTION OF A THIRD MOLAR AT THE USER FACILITY, THE DRILL WAS OVERHEATING. NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE EXTRACTION OF A THIRD MOLAR AT THE USER FACILITY, THE DRILL WAS OVERHEATING. NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
46063 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1