ADM BALL IMPACTOR TIP
Report
- Report Number
- 0002249697-2013-00547
- Event Type
- Malfunction
- Date Received
- February 4, 2013
- Date of Event
- January 17, 2013
- Report Date
- January 17, 2013
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- NURSE
Narratives
THE ROOT CAUSE WAS DETERMINED TO BE USER RELATED. THE VISUAL INSPECTION AND MATERIAL ANALYSIS CONCLUDED THAT THE IMPACTOR TIP BROKE THROUGH THE THREADS AT APPROXIMATELY THE THIRD THREAD FROM THE MOUNTING SURFACE. MULTIPLE FRACTURE INITIATION SITES WERE OBSERVED ON THE FRACTURE SURFACE. THE FRACTURE PROGRESSED IN MULTIPLE DIRECTIONS. THE PART FAILED FROM MULTIPLE IMPACTS AS EVIDENCED BY THE ARREST LINES VISIBLE ON THE FRACTURE SURFACE. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE FRACTURE SURFACES EXAMINED. DEVICE HISTORY RECORDS FOR THE SPECIFIC LOT INDICATED THAT THE DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. A REVIEW OF THE COMPLAINT HISTORY DATABASE SHOWS THAT THERE HAVE BEEN NO SIMILAR REPORTED EVENTS FOR THE SUBJECT LOT CODE.
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
THE NURSE REPORTED THAT THE BALL TIP GOT BROKEN DURING THE OPERATION. ANOTHER TIP WAS USED TO COMPLETE THE OPERATION. OPERATION GOT LONGER ONLY FEW MINUTES.
THE NURSE REPORTED THAT THE BALL TIP GOT BROKEN DURING THE OPERATION. ANOTHER TIP WAS USED TO COMPLETE THE OPERATION. OPERATION GOT LONGER ONLY FEW MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 46347 | ADM BALL IMPACTOR TIP | INSTRUMENT | LXH | STRYKER ORTHOPAEDICS-MAHWAH | G3190596 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |