FDA Adverse Event Malfunction Summary report: N

ADM BALL IMPACTOR TIP

MDR report key: 2941859 · Received February 4, 2013

Report

Report Number
0002249697-2013-00547
Event Type
Malfunction
Date Received
February 4, 2013
Date of Event
January 17, 2013
Report Date
January 17, 2013
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE ROOT CAUSE WAS DETERMINED TO BE USER RELATED. THE VISUAL INSPECTION AND MATERIAL ANALYSIS CONCLUDED THAT THE IMPACTOR TIP BROKE THROUGH THE THREADS AT APPROXIMATELY THE THIRD THREAD FROM THE MOUNTING SURFACE. MULTIPLE FRACTURE INITIATION SITES WERE OBSERVED ON THE FRACTURE SURFACE. THE FRACTURE PROGRESSED IN MULTIPLE DIRECTIONS. THE PART FAILED FROM MULTIPLE IMPACTS AS EVIDENCED BY THE ARREST LINES VISIBLE ON THE FRACTURE SURFACE. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE FRACTURE SURFACES EXAMINED. DEVICE HISTORY RECORDS FOR THE SPECIFIC LOT INDICATED THAT THE DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. A REVIEW OF THE COMPLAINT HISTORY DATABASE SHOWS THAT THERE HAVE BEEN NO SIMILAR REPORTED EVENTS FOR THE SUBJECT LOT CODE.

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

THE NURSE REPORTED THAT THE BALL TIP GOT BROKEN DURING THE OPERATION. ANOTHER TIP WAS USED TO COMPLETE THE OPERATION. OPERATION GOT LONGER ONLY FEW MINUTES.

Description of Event or Problem · 1

THE NURSE REPORTED THAT THE BALL TIP GOT BROKEN DURING THE OPERATION. ANOTHER TIP WAS USED TO COMPLETE THE OPERATION. OPERATION GOT LONGER ONLY FEW MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
46347 ADM BALL IMPACTOR TIP INSTRUMENT LXH STRYKER ORTHOPAEDICS-MAHWAH G3190596

Patients

Seq Age Sex Outcome Treatment
1 Other