AUTOPULSE NIMH BATTERY
Report
- Report Number
- 3003793491-2012-00544
- Event Type
- Malfunction
- Date Received
- December 27, 2012
- Date of Event
- October 12, 2012
- Report Date
- October 23, 2012
- Manufacturer
- ZOLL CIRCULATION, INC.
- Product Code
- DRM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE AUTOPULSE NIMH BATTERY (S/N (B)(4)) INVOLVED IN THE EVENT WAS RETURNED TO ZOLL CIRCULATION ON (B)(4) 2012 AND WAS ANALYZED. THE CUSTOMER IS NOT FOLLOWING PROPER BATTERY MANAGEMENT PROCEDURES OF DAILY SYSTEM CHECKS AND BATTERY SWAP. LOW VOLTAGE BATTERY HAVE BEEN USED REPEATEDLY. ON (B)(4) 2012, THIS BATTERY WAS USED REPEATEDLY FOR DEPLOYMENT WHICH WAS NOT FULLY CHARGED WHICH MAY HAVE CAUSED THE REPORTED COMPLAINT OF "USER ADVISORY 17." AFTER THE BATTERY WAS CHARGED AND TEST CYCLED, IT PASSED TESTING. NO ADVERSE EVENT REPORTED.
IT WAS REPORTED THAT THE DEVICE STOPPED COMPRESSION AFTER 10 MINUTES. CUSTOMER WILL RETURN THE DEVICE AND TWO BATTERIES (S/N (B)(4)). CUSTOMER ALSO REQUESTED INVESTIGATION OF THIS FAILURE AND REPAIR OF USER ADVISORY 17 (MAX MOTOR ON TIME EXCEEDED DURING ACTIVE OPERATION).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AUTOPULSE NIMH BATTERY | NIMH BATTERY | DRM | ZOLL CIRCULATION, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |