LINOX SD 65/16
Report
- Report Number
- 1028232-2012-03234
- Event Type
- Malfunction
- Date Received
- December 21, 2012
- Date of Event
- October 9, 2012
- Report Date
- December 6, 2012
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- LWS
- PMA / PMN Number
- P980023
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
OUS MDR.
THE LEAD UNDER COMPLAINT WAS EXTENSIVELY ANALYZED, INCLUDING A VISUAL AND AN ELECTRICAL INSPECTION. THE ANALYSIS REVEALED SEVERE ABRASION MARKS ALONG THE LEAD BODY. IN PARTICULAR, AT THE DISTAL PART OF THE LEAD THE INSULATION WAS RUBBED THROUGH. THIS INSULATION DAMAGE CAN BE CONSIDERED AS THE ROOT CAUSE FOR THE OBSERVED OVERSENSING ISSUES MENTIONED IN THE COMPLAINT DESCRIPTION. BASED ON THE CHARACTERISTICS OF THIS DAMAGE, IT IS REASONABLE TO ASSUME THAT THE LEAD HAD BEEN SUBJECT TO SEVERE MECHANICAL STRESS IN THE IMPLANTED STATE. DIAGNOSTIC IMAGES CLARIFYING THIS ASSUMPTION WERE NOT AVAILABLE FOR ANALYSIS. THE CUTTINGS OF THE INSULATION RESULTED MOST LIKELY FROM THE EXPLANTATION PROCEDURE. THE ANALYSIS DID NOT REVEAL ANY SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.
OUS MDR - AFTER AN IMPLANT DURATION OF ABOUT 39 MONTHS OVERSENSING WAS REPORTED. NO ADVERSE PT SIDE EFFECTS HAVE BEEN REPORTED. THE LEAD WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LINOX SD 65/16 | ICD LEAD | LWS | BIOTRONIK SE & CO. KG | 350053 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Hospitalization |