FDA Adverse Event Malfunction Summary report: N

LINOX SD 65/16

MDR report key: 2941709 · Received December 21, 2012

Report

Report Number
1028232-2012-03234
Event Type
Malfunction
Date Received
December 21, 2012
Date of Event
October 9, 2012
Report Date
December 6, 2012
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
LWS
PMA / PMN Number
P980023
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

OUS MDR.

Additional Manufacturer Narrative · 1

THE LEAD UNDER COMPLAINT WAS EXTENSIVELY ANALYZED, INCLUDING A VISUAL AND AN ELECTRICAL INSPECTION. THE ANALYSIS REVEALED SEVERE ABRASION MARKS ALONG THE LEAD BODY. IN PARTICULAR, AT THE DISTAL PART OF THE LEAD THE INSULATION WAS RUBBED THROUGH. THIS INSULATION DAMAGE CAN BE CONSIDERED AS THE ROOT CAUSE FOR THE OBSERVED OVERSENSING ISSUES MENTIONED IN THE COMPLAINT DESCRIPTION. BASED ON THE CHARACTERISTICS OF THIS DAMAGE, IT IS REASONABLE TO ASSUME THAT THE LEAD HAD BEEN SUBJECT TO SEVERE MECHANICAL STRESS IN THE IMPLANTED STATE. DIAGNOSTIC IMAGES CLARIFYING THIS ASSUMPTION WERE NOT AVAILABLE FOR ANALYSIS. THE CUTTINGS OF THE INSULATION RESULTED MOST LIKELY FROM THE EXPLANTATION PROCEDURE. THE ANALYSIS DID NOT REVEAL ANY SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

OUS MDR - AFTER AN IMPLANT DURATION OF ABOUT 39 MONTHS OVERSENSING WAS REPORTED. NO ADVERSE PT SIDE EFFECTS HAVE BEEN REPORTED. THE LEAD WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LINOX SD 65/16 ICD LEAD LWS BIOTRONIK SE & CO. KG 350053

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization