LIFEVEST WCD 4000 SYSTEM
MDR report key: 2941681
·
Received December 21, 2012
Report
- Report Number
- 3008642652-2012-03317
- Event Type
- Malfunction
- Date Received
- December 21, 2012
- Date of Event
- November 29, 2012
- Report Date
- December 14, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESPONSE BUTTONS NOT WORKING) WAS CONFIRMED. UPON INVESTIGATION THE REAR RESPONSE BUTTON WAS NOT FUNCTIONAL. THE CAUSE FOR THE NON-FUNCTIONAL RESPONSE BUTTON WAS A DETACHED RESPONSE BUTTON CABLE. THE ROOT CAUSE FOR THE DETACHED CABLE COULD NOT BE POSITIVELY IDENTIFIED, BUT THE CABLE LIKELY DISCONNECTED AS A RESULT OF THE MONITOR IMPACTING A HARD SURFACE. NO ADVERSE EVENT RESULTED FROM THE DETACHED RESPONSE BUTTON CABLE. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR'S RESPONSE BUTTONS WERE NOT WORKING. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.