FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2941672 · Received December 21, 2012

Report

Report Number
3008642652-2012-03307
Event Type
Malfunction
Date Received
December 21, 2012
Date of Event
October 25, 2012
Report Date
December 14, 2012
Manufacturer
ZOLL LIFECOR CORP.
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (EXPOSED WIRE) HAS BEEN CONFIRMED. UPON INVESTIGATION THE CABLE CONNECTING THE DISTRIBUTION NODE (DN) AND REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, EXPOSING WIRES. THE ROOT CAUSE FOR THE STRAINED CABLE WAS EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE STRAINED CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

THE WIFE OF A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT A WIRE WAS EXPOSED ON THE PATIENT'S ELECTRODE BELT. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORP. WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 76 YR