FDA Adverse Event Injury Summary report: N

QUADRA H FEMORAL STEM

MDR report key: 2941611 · Received January 7, 2013

Report

Report Number
3005180920-2012-00098
Event Type
Injury
Date Received
January 7, 2013
Date of Event
November 24, 2012
Report Date
January 7, 2013
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
PMA / PMN Number
K082792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DOCUMENT REVIEW: QUADRA H CEMENTLESS FEMORAL STEM SIZE 6 STD - REF. 01.12.026 / LOT 082944 (32 STEMS PRODUCED): ALL PARAMETERS WERE FOUND TO BE IN ACCORDANCE WITH THE SPECIFICATIONS VALID AT THE TIME OF MANUFACTURING. THE WASHING CYCLE FOR METAL COMPONENTS WAS RUN ACCORDING TO SPECIFICATIONS AND NO ALARMS OR DEVIATION OCCURRED DURING OPERATION. THE STERILIZATION CYCLE WAS PERFORMED ACCORDING TO SPECIFICATIONS VALID AT THE TIME OF PRODUCTION. THE 24 STEMS BELONGING TO THIS LOT HAVE BEEN ALREADY IMPLANTED AND NO OTHER INCIDENTS HAVE BEEN REPORTED UP TO NOW. CERAMIC FEMORAL HEAD MANUFACTURED BY(B)(4) AND ONLY DISTRIBUTED BY MEDACTA: WE BOUGHT 98 HEADS OF THE LOT INVOLVED. THE 80 HEADS HAVE BEEN ALREADY IMPLANTED WITHOUT ANY OTHER SIMILAR CASE. VERSAFITCUP CC HIGHCROSS PE LINER (K103352): REF. 01.26.364HCT - LOT 082309 (31 INSERTS PRODUCED): NO ANOMALIES FOUND IN THE DOCUMENT REVIEW, AS FOR THE STEM. THE 29 LINERS OF THIS LOT HAVE BEEN ALREADY IMPLANTED WITHOUT ANY OTHER SIMILAR INCIDENT. VERSAFITCUP CC LIGHT SHELL - NOT MARKETED IN THE USA - LOT 082891 (20 CUPS PRODUCED): NO ANOMALIES FOUND IN THE DOCUMENT REVIEW, AS FOR THE STEM. THE 19 CUPS OF THIS LOT HAVE BEEN ALREADY IMPLANTED WITHOUT ANY OTHER SIMILAR INCIDENT. FROM THE DATA COLLECTED, THERE IS NO EVIDENCE THAT THE INFECTION OCCURRED IS DEVICE RELATED.

Description of Event or Problem · 1

REVISION SURGERY DUE TO INFECTION 3 YEARS POST OP. ALL THE COMPONENTS REVISED. WE WERE INFORMED ON (B)(6) 2012 ONLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
7538 QUADRA H FEMORAL STEM FEMORAL STEM SIZE 6 STD CEMENTLESS JDI MEDACTA INTERNATIONAL SA 082944

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention IN USA. HX PE LINER CODE 01.26.3648HCT LOT 082309| (K103352)| CERAMIC BALL HEAD ANDACETABULAR CUP NOT MARKETED