FLOGARD
Report
- Report Number
- 1416980-2013-02766
- Event Type
- Malfunction
- Date Received
- February 4, 2013
- Date of Event
- November 19, 2012
- Report Date
- November 19, 2012
- Manufacturer
- SHARP CORPORATION OSAKA JAPAN
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). THE DEVICE HAS BEEN RECEIVED, AND THE CONDITION WAS CONFIRMED BY SERVICE.¿THIS IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE OF THE DAMAGED POWER CORD WAS DETERMINED TO BE PHYSICAL DAMAGE TO THE INSULATING TAPE. TO CORRECT THE CONDITION THE POWER CORD WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL INFORMATION IS RECEIVED A FOLLOW UP MDR WILL BE SENT.
DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A DAMAGED POWER CORD. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS WAS NOT REPORTED BY THE CUSTOMER. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 46232 | FLOGARD | PUMP, INFUSION | FRN | SHARP CORPORATION OSAKA JAPAN |