FDA Adverse Event Malfunction Summary report: N

PLUM A+ UK WITH 3PIN

MDR report key: 2941232 · Received January 18, 2013

Report

Report Number
9615050-2013-00105
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
December 28, 2012
Report Date
December 28, 2012
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K070398
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THE DEVICE WAS RETURNED TO THE CLINICAL ENGINEERING DEPARTMENT WITH AN UNSIGNED NOTED THAT STATED, "FAULTY". NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT ALARM WHEN A DISTAL OCCLUSION WAS PRESENT. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
28301 PLUM A+ UK WITH 3PIN 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK PLUM A+ SOFTWARE MODULE, LIST #12097,| SN (B)(4)