FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 2941217 · Received January 18, 2013

Report

Report Number
2937094-2013-00108
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
October 9, 2012
Report Date
October 24, 2012
Manufacturer
AMERICAN MEDICAL SYSTEMS
Product Code
GEX
PMA / PMN Number
K100746
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPONENT CODES: FIBER AND CAP REFER TO THE RESULTS CODE THERMAL PROBLEM. FIBER ANALYSIS: THE FIBER CAP EXHIBITS A RADIAL FRACTURE AT THE OUTPUT WINDOW OF THE CAP; THE FIBER IS FRACTURED DISTAL TO THE FIBER/CAP FUSION ZONE. THE FIBER PROXIMAL TO FRACTURE ROTATES INDEPENDENTLY OF METAL CAP. THE METAL CAP SHOWS SIGNS OF BURNT ON DETRITUS AND THE GLASS CAP SHOWS SIGNS OF DEVITRIFICATION. THE FIBER/CAP CONDITIONS WOULD RESULT IN FORWARD FIRING. THE MOST PROBABLE ROOT CAUSE THAT CONTRIBUTED TO THIS FAILURE WAS SUSPECTED TO BE RELATED TO LOCALIZED HEAT ACCUMULATION/USER HANDLING DUE TO ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE PROCEDURE, THERE WAS DECREASED VAPORIZATION AT 83,786 JOULES. THE CASE WAS COMPLETED WITH A SECOND FIBER. NO INJURY TO PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
28400 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS 10-2400 202B

Patients

Seq Age Sex Outcome Treatment
1 GREENLIGHT XPS LASER SYSTEM AND ACCESSORIES