FDA Adverse Event
Malfunction
Summary report: N
6800
MDR report key: 2941141
·
Received January 18, 2013
Report
- Report Number
- 1720753-2013-00842
- Event Type
- Malfunction
- Date Received
- January 18, 2013
- Date of Event
- January 10, 2013
- Report Date
- January 18, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE SINGLE BOARD COMPUTER AND ADJUSTED THE BRIGHTNESS AND CONTRAST CONFIGURATION SETTINGS. THE SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE IMAGES WERE WHITE. THIS WILL CAUSE THE SYSTEM TO BE UNUSABLE DUE TO THE INABILITY TO SEE THE LIVE IMAGE. THERE IS NO REPORT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 28256 | 6800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 6800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |