FDA Adverse Event Malfunction Summary report: N

TRILOGY 202

MDR report key: 2941134 · Received January 16, 2013

Report

Report Number
2518422-2013-00067
Event Type
Malfunction
Date Received
January 16, 2013
Date of Event
December 28, 2012
Report Date
December 28, 2012
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K093905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A STEP DURING TESTING. THE DEVICE'S OXYGEN BLENDING MODULE BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
25130 TRILOGY 202 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1040007

Patients

Seq Age Sex Outcome Treatment
1