FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 2941121
·
Received January 16, 2013
Report
- Report Number
- 2518422-2013-00082
- Event Type
- Malfunction
- Date Received
- January 16, 2013
- Date of Event
- January 4, 2013
- Report Date
- January 4, 2013
- Manufacturer
- RESPIRONICS INC.
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, "SERVICE REQUIRED" CODES WERE FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD AND POWER MANAGEMENT BOARD WERE REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
A VENTILATOR WAS RETURNED TO THE MFR FOR SERVICE. THE DEVICE WAS NOT IN PT USE.